TECHNICAL INFORMATION | |
Measuring Principle | The QCM-I unit is a high sensitivity mass sensor which measures |
Measurement Channels | 1-4 |
Frequency Range | 1-80 MHz,up to the 13th overtone for a 5 MHz Crystal |
Working Temperature | 4° C to 80° C, controlled manually or via software, |
Sample Loading | Sample injection with external syringe or peristaltic pump unit |
Flow Cell Volume | ~ 40 µl (typical with Ø14 mm crystals) |
Typical Flow Rates | 0.01µl/min - 35ml/min with external syringe pump, Syringe size 1-60 ml |
Sample Volume | 1 to 2000 µl (depends on installed loop volume) |
Wetted parts | PTFE, PEEK, SS, VITON or Kalrez |
MEASUREMENT SENSITIVITY | |
Standard Resonance Frequency sensitivity | ~2 x 10-1 Hz |
Standard Dissipation Sensitivity in Liquid | ~ 1 x 10-7 |
Mass Sensitivity in Liquid | ≤ 1 ng/cm2 |
SENSORS | |
Sensor Crystal | 5 MHz, AT-cut, modular sensor holder accommodates either 14 mm or 1” inch diameter crystals See more details inQCM sensor datasheets. |
SOFTWARE | |
BioSense | Universal software platform for OWLS, QCM & EC measurements See more details in BioSense datasheet. |
Import/Export of the measured data | Between separate BioSense software and export to third party software Excel, JPG, BPM, WMF etc. |
PC Requirement | USB 2.0, works with MS Windows® 7 or 8 or 10 |
ELECTRICAL | |
Supply Voltage | DC power supply with universal input voltage from 100V to 240V AC & 50 to 60 Hz. |